Microluminescent zero instrument
Category:
Tel:
Product Details
Microlight Zero-Position Instrument
Applications and Main Features
The microlight zero-position instrument adopts a Newtonian reflection system and is mainly used to detect the zero-position drift and optical axis consistency of white light and microlight products.
With an additional resolution board, it can also test the resolution image quality indicators of white light/microlight products. 。 Additional With an additional infrared crosshair reticle and infrared light source, it can also measure the zero-position drift of infrared thermal imaging products.
Main Technical Indicators:
Effective aperture of objective lens: φ300mm;
Focal length of objective lens: 2000mm;
Central obscuration: φ85mm;
Surface type: Parabolic reflecting mirror;
Surface accuracy: RMS better than λ/40, λ=632.8nm;
The zero-position drift is measured roughly through the reticle grid value, with the minimum grid value being 30″, and precisely measured through electronic control, with the minimum grid value being better than 1″.