Multi-axis calibration test system
Category:
Tel:
Product Details
The multi-optical axis calibration test system is used for the optical axis calibration and optical parameter testing of integrated optoelectronic products, such as integrated optoelectronic products including infrared aiming components, white light television aiming components, laser rangefinder components, and multi-aperture high-energy laser emission components, and analyzes the quality of the laser beam and measures the laser output power.
The system uses a Φ500mm off-axis parabolic system as the main test optical path, using a giant wedge mirror for beam splitting and attenuation, high-precision plane mirrors and electrically controlled guide rails for optical path switching, and a wide-spectrum CCD camera to collect the measurement data, and finally obtain various parameters through computer software processing.
Functions
1 ) Achieve laser power and beam quality testing ;
2 ) Achieve single-beam laser collimation testing ;
3 ) Achieve multi-beam optical axis parallelism testing ;
4 ) Achieve parallelism of multispectral visual axes testing ;
5 ) Achieve parallelism between the visual axis and the emitted laser optical axis testing ;
6 ) System self-check function.
Technical Indicators
1 ) Single-beam laser beam collimation test error: <0.01mrad ;
2 ) Test error of parallelism between laser beams: <0.01mrad ;
3 ) Test error of parallelism between laser axes and visual axes: <0.01mrad ;
4 ) Test error of spatial arrangement of laser optical axis array spots: <0.1% 。